FP-6500/6600 Flourescence Spectrometer

Contact Us

FP-6500/6600 Flourescence Spectrometer

The FP-6500 is the research-grade spectrofluorometer with a minimum 1 nm resolution and a highest sensitivity S/N ratio 200 or greater for the Raman band of water.

The FP-6500 is especially designed to keep in mind in order to meet requirements in biochemistry as kinetics, stopped-flow, titration, or anisotropic measurements.

The FP-6500 is designed to keep GLP/GMP compliance in mind. Built-in low pressure Mercury lamp and UV cut filter enable to the GLP/GMP compliance with validation software which is available as standard.

The FP-6600 is a factory optional model of the FP-6500 which can be covered emission measurement of NIR region up to 1010 nm by using the holographic grating with 900 grooves/mm and R2658P photomultiplier tube for the emission monochromator.

Features

Large Throughput Optics
The sensitivity is one of most important factor in performance of Fluorescence Spectrophotometer. The FP-6500 is proud of S/N ratio for Raman band of water at 350 nm is greater than 200:1 (SBW: 5 nm on both EX and EM, response: 2 sec., peak-to-peak). It corresponds to greater than 2400:1 at same conditions (SBW: 10 nm on both EX and EM, response: 2 sec., RMS) as competitors.

Meeting the Requirements for Biochemistry
A trigger switch for starting measurement is available on the main unit as standard, and also enables to control by an operation on the PC. This function is useful to perform Kinetics analysis. Optional time course software accomplished fast response of 0.1 millisecond, and is very effective for stopped-flow measurement.

Half slit height(L5 and L10 nm) functions for spectral bandwidths enables to meet the requirements on sample measurements using micro cell.

High Quality Optical System
>The optical system utilizes high quality components including a DC-powered 150W Xenon lamp (in a sealed housing), holographic concave grating monochromators, fast EX and EM direct scanning drive, variable slits and horizontal sample illumination. Especially, minimum spectral bandwidth of 1 nm on both EX and Em monochromators can get high resolution data.

Schematic of the FP-6000 series (Image rotation optics)

Horizonal Slit Image
The sample illumination optics incorporate an all reflective image rotation optical system to obtain high emission yields. Rotating the slit image of the monochromator formed on the sample cell by 90 deg., the observed volume of the sample illumination is enlarged by the slits length-to-width ratio, resulting in higher emission intensity providing an increase in sensitivity.

GLP/GMP Compliance
The FP-6500 is designed to keep GLP/GMP compliance in mind. Built-in low pressure Mercury lamp and UV cut filter enables to the GLP/GMP compliance with validation software which is available as standard. Validation program is based on JAIMAS, and it is not necessary to carry out traceability on jig.

Phosphorescence Measurement
The FP-6500 is supplied Phosphorescence spectrum measurement program and Phosphorescence lifetime measurement program as standard.

Compact Design
The compact optical bench is housed in a rugged case measuring only 550 (W) x 540 (D) x 300 (H) mm.


Spectra Manager ™

JASCO is the first instrument manufacturer to offer a powerful 32 bit Windows software platform for control of it's full range of spectroscopy instrumentation as well as complete analysis and display of any kind of spectral data. Spectra Manager tm is an innovative approach to a cross-platform package setting a new standard for ease of use and allowing users to work seamlessly between several different types of instrumentation.